HD 9009TRR | transducer for temperature and humidity

Product.Nr.: HD9009TRR

Manufacturer: Senseca (ehemals DeltaOhm)

EUR 368,00
incl. 19 % VAT

  • delivery time bzw. Fertigungszeit variiert - bitte erfragen!

product description

Dual, meteorological microprocessor transducer for humidity and temperature,
4 mA to 20 mA output

 

General information:

HD9009TRR is a single block microprocessor transducer for humidity and temperature, temperature is configurable. The HD9009TRR is a transducer with a 0 V to 1 V standard voltage output (other outputs available on demand) and 5 V DC to 35 V DC power supply. The sensors are mounted at the end of a plastic tube: a capacitive humidity sensor and a platinum temperature sensor (100 Ω @0°C).

The instrument can be reprogrammed by means of a key, and no jumper or potentiometer actions are required. The humidity input can be recalibrated by using two saturated solutions: the first one at 75 %, the second one at 33 %; the elative humidity range 0 % to 100 % r.F. r is fixed, 0 V DC correspond to 0 % r.F., 1 V DC equal 100 % r.F.

The temperature standard configuration is –40°C to +80°C, corresponding to 0 V DC to 1 V DC. The user can configure the temperature output in ranges different from the standard one by means of a Pt100 simulator or of a set of fixed resistances, provided that it is included within the -40°C to +80°C range and shows a minimum amplitude of 25°C. Two LEDs give alarm indications (temperature exceeding set range, sensor breakage or shortcircuit) and help the operator when programming.

An out-of-standard temperature operating range can be requested when placing the order.

Important warning: the probes work in the -40°C to +80°C temperature range. Outside of this range data are not correct; electronics are designed to operate within this range.

Downloads